  • How it Works: Measuring is based on the principle, that when mixing gases with different thermal conductivities their mixture has a thermal conductivity dependent on the concentration of its constituents. Thereby portions of individual gases can be determined.
  • Ease of Use: Just connect the device to the process stream, switch on. After 30 min (heat up time) the device is ready for use.
  • The measurement are based on the thermal conductivity. Please keep in mind that there some restrictions.

Application:

  • Precise analysis of (quasi) binary gas mixtures
  • Sehr hohe Präzision z.B. H2 in N2: 30ppm
  • H2 in Luft, He in Ar, C02 in N2 u. v. m.
  • Measurement according to the thermal conductivity principle
  • Solution of up to 99 tasks integrated

Communication:

  • Key-controlled menu navigation with display
  • Current output: 1 x 0 / 4-20mA
  • Voltage output: 2 x 0-10V
  • Control and signal reading via RS232
  • 3 limit and alarm relays
  • Voltage input

Construction:

  • Dimensions (W x H x P in mm): (145 x 80 x 85)
  • Weight: 1800 g
  • Protection class: IP65
  • Connection: 6mm pipe
  • Pressure-proof: 20bar absolute
General: TCD3000
Measurement process Thermal conductivity
Sampling Analyzer extractive
Connection 6mm pipe
Dimensions (WxHxD in mm) 145 x 80 x 85
Protection class IP 65
Pressure-proof to 20 bar absolute
Power supply 18V to 36V DC / 700mA
Ambient temperature range -20°C to 50°C
Weight to 1800g
T90 time at 60l/h < 1sec
Communication:
RS232 Yes
Current output 1x, 0/4-20 mA
Voltage output 2x, 0-10V
Display Yes
Relays 3x
Service program SetApp 2.0 Yes
Calibration on site Yes
Options:
Flow measurement Yes
Flow monitor Yes
Multi Gas Mode Yes, additional up to 99 gas pairs
Protection against corrosion Yes
Protection against condensate and dust Yes
Suitable for inflammable gases Yes
Additional Modules Compatible Yes (O2)
Moisture measurement Yes
Cross-sensitivity compensation Yes, external signal required
Specifications gas analysis:
Noise < 1% of smallest range
Drift at zero point per week < 2% of smallest range
Repeatability < 1% of smallest range
Non-linearity < 1% of range
Measuring error with ambient temperature change per 10°K < 1% of smallest range
Flow influence between 60l/h and 90l/h per 10l/h < 1% of smallest range
Fault with measurement gas change (Pabs > 800 hPa) per 10 hPa < 1% of smallest range

Downloads

About function:

Measuring is based on the principle, that when mixing gases with different thermal conductivities their mixture has a thermal conductivity dependent on the concentration of its constituents. Thereby portions of individual gases can be determined.

About calibration:

We recommend to do a calibration at one point with a pure gas or bottle gas. After the gas is applied to the device wait until the concentration is stable. The calibration process can be started via the keys on the front of the analyzer or via RS232 command. Then it takes ten seconds for the actual calibration.

We recommend a calibration, resp. a check of the calibration if one of the following criteria is met:

  • After bringing the device into service
  • On a regular cycle, depending on the precision aimed for.

    To find out the appropriate time between calibrations, we recommend to begin with a more frequent check of the calibration.

    The time between calibrations can range between:

    • several month for a measuring task in the Vol.% range
    • days to weeks for a measuring task in the sub-Vol.% range
    • directly before every measurement if highest accuracy is needed
  • When the situation of the measurement regarding pressure, temperature or gas flow changes

About measuring ranges:

    We offer 0-100 Vol.% H2 in N2 (and many other background as Ar, CO2, O2, CH4…)

  • The smallest range for low H2 concentration is 0-0.5Vol.% H2 in N2.
  • The smallest range for high H2 concentration is 98-100Vol.% H2 in N2.

    Helium is similar with the ranges:

  • The smallest range for low He concentration is 0-0.8Vol.% He in N2.
  • The smallest range for high He concentration is 97-100Vol.% He in N2.

Any range in between is available.

About operation:

The analyzer need pneumatic connection to the sample gas. It is designed for wall mounting.

The analyzer comes with cables with open ends. Connection to 24V is necessary for operation.

Connection of 4-20mA output, relays and RS232 are available and can be connected as required.

Choosing the right solution:

If you are not sure whether this unit is suitable for your requirements, please contact us for product advice.
In download area you can find a measuring task form, which supports you find out specification of your application.

If other gas pairs are required, please contact our sales team.

CONTACT US

Please do not hesitate to contact us if you would like a quote or have a question. You can reach us via the contact form and via the e-mail address below. Before making an offer, we clarify the requirements with a form for the measuring task. This way we want to make sure that you will receive the device from us that best fits your measuring task.

Contact Info

phone : +49 (0)69 247544980

e-mail:   info@archigas.de

When submitting the data entered in the contact form by clicking on the button below, you declare that you have read the privacy policy and that you agree that the data provided will be collected and stored electronically.
The data is used only strictly earmarked for processing and answering your request. By submitting the contact form you agree to the processing. You may revoke your consent at any time with future effect. In case of cancellation your data will be deleted immediately.